All Collections
Monitor Connections
Mindray ePM 10a and ePM 12 Settings
Mindray ePM 10a and ePM 12 Settings

How to configure Mindray ePM series for use with Xchart

Henrik Joreteg avatar
Written by Henrik Joreteg
Updated over a week ago
  1. Select "Main Menu"

  2. Swipe left to get to "Maintenance"

  3. Enter password (uppercase): MIN332
    If that password doesn't work try: 332888
    tip: you can click the little "eye" icon to make sure you typed it correctly

  4. Now find the little ">>" icon at the top right:

  5. Go to the "Other" tab and select "Parameter Output Setup" to exactly this:

    1. Baud Rate: 38400

    2. Parity Mode: None

    3. Data Bits: 8

    4. Stop Bits: 1

  6. Next, still in the "Maintenance" window, use the arrows again to locate the "Factory Maintenance" tab and then find "Setup" in the side tabs that appear:
    ​

  7. Important NOT all monitors have this if there is a "BeneLink" switch as shown below, make sure it is enabled. If not CONTINUE TO STEP 8.
    ​

  8. Next click "Update ID module" and set the setting exactly as follows:

    1. BeneLink Module Port: 1

    2. ID: 6000a000

    3. Click "Confirm"

    4. Make sure it says "Upgrade Successful"

  9. For good measure, restart the monitor completely.

At this point you should be all set to use this monitor with Xchart.
​
If you prefer to see this all in video form, see video below, just beware that Step 7 up above, was added after this video was produced. So if there's a BeneLink switch make sure it's enabled: