Monitor settings for Mindray VS9 if using Mindray DIAP cable

Not relevant if using Xlink

Henrik Joreteg avatar
Written by Henrik Joreteg
Updated over a year ago

Tap upper left corner and make sure it it set to "Continuous Profile"
​

Navigate to:
​

  1. Main Menu

  2. Maintenance

  3. Password: MIN888

  4. Use << and >> symbols in upper right corner to find and select Other tab

  5. Click Paremeter Output Setup set the following:

    1. Baud Rate: 38400

    2. Parity Mode: None

    3. Data Bits: 8

    4. Stop Bits: 1

Parameter Outpue Setup should look like this:


​Connect it physically as follows: