All Collections
Monitor Connections
Mindray DPM 6 Configuration and Connection
Mindray DPM 6 Configuration and Connection

How to connect the Mindray DPM 6 to Xchart using the Xlink adapter

Henrik Joreteg avatar
Written by Henrik Joreteg
Updated over a week ago

To connect the Mindray DPM 6 you'll need an Xlink box from us, here at Xchart. You can get one, pre-configured for the DPM series, by emailing us at: [email protected]
โ€‹

Step #1: Configure the monitor (only need to do this once)

There is only one setting you need to change on your monitor and you get to it as follows:

  1. Menu

  2. Maintenance

  3. User Maintenance

  4. Enter password: 888888

  5. Network Setup.

  6. Enter this exact IP Address: 192.168.0.201

Step #2: Connect your Xlink kit to the back of the monitor as shown here:
โ€‹

Step #3: Open case in Xchart and select "Mindray HL7" when connecting your monitor

If you need help, reach out to: [email protected]

Did this answer your question?