All Collections
Monitor Connections
Configure Mindray ePM 10, ePM 12 for Xlink adapter
Configure Mindray ePM 10, ePM 12 for Xlink adapter
Henrik Joreteg avatar
Written by Henrik Joreteg
Updated over a week ago

If you prefer the text instructions see below the video:
โ€‹

Navigate the menus as follows:

  1. Main Menu

  2. Maintenance

  3. User Maintenance

  4. Password: MIN888

  5. Use ">>" sign in upper right to get to the "Network Setup" tab.

  6. Select the left tab marked "LAN1 IP" make it say exactly this:

    1. Use the Following Address:

    2. IP Address: 192.168.0.5

    3. Subnet Mask: 255.255.255.0

    4. Gateway: 192.168.0.1

  7. Select HL7 Configuration tab on the left. Set it to:

    1. Server Address: 192.168.0.7

    2. Port: 5000

    3. Data Inteval: 10 seconds

    4. Enable the "Send Data" switch

    5. If your account has the waveforms feature enable the "Send Waveforms" switch.

  8. Hit the "X" to close the menu

  9. Restart the monitor.


โ€‹

Did this answer your question?